I need to know:
1. Can the Micro GC has data logging ability and pre-loaded with analysis program i.e it need not require a Notebook PC to hook on to the GC during analysis?
2. Beside Hewlett-Packard who else supply Micro GC and where can I get the specification?
![]()
![]()
![]()
![]()
By Anonymous on Friday, August 11, 2000 - 11:44 am:
1)- No ie: yes it does www.chem.agilent.com/scripts/generic.asp?lpage=245
2)- Search @ Yahoo,Google,Hotbot,Altavista, etc.
![]()
![]()
![]()
![]()
By Anonymous on Wednesday, January 24, 2001 - 08:38 am:
Varian also sells a micro GC.
![]()
![]()
![]()
![]()
By Anonymous on Saturday, March 3, 2001 - 09:15 am:
can micro-GC be connected to thermal analyzer, so the inlet to the GC will be the gases evolved from thermal analyzer.
Thank you
![]()
![]()
![]()
![]()
By formic on Friday, April 27, 2001 - 01:14 pm:
Anonymous-
A Micro GC can be connected to almost any type of equipment that produces gas. We have one hooked up to a multi port valve and are able to analyze input gasses and output process gasses from an autoclave used as a reactor for FT gas to liquids technology. CAUTION- on HP Quad Micro GC's, don't exceed 30psi into the injector and make sure the sample contains as little water as possible. This is especially critical if you have a Mol Sieve column on one of the channels.
Hope this helps some.