I am trying to find a single capillary column (prefer 0.32um) to analyze natural gas. I would like to separate N2, CH4, CO2, & C2-C6. The goal is to reduce the run time to seven minutes or less. We currently use micropacked columns for the analysis.
Any recommendations or suggestions would be helpful.
![]()
![]()
![]()
![]()
By Jason Ellis on Tuesday, April 16, 2002 - 08:20 am:
Porous polymer PLOT columns work well for natural gas analysis. I have some chromatograms that I could refer you to. If you're interested, please contact me directly.
![]()
![]()
![]()
![]()
By Rodney George on Wednesday, April 17, 2002 - 08:48 pm:
One can do this analysis in less than 6 minutes with micropacked columns and get excellent reproducibility and separation with an isothermal oven (no time delay for cool down).