I haven't got a much expirience in chromatohraphy and I would be very grateful for any help. I need a method for phthalates determination in alcoholilc beverages (plum brandy).We have a GC with capillary column. I'm interesting in quck sample preparation without concetration.
Gratefully,
Ana
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, April 30, 2002 - 08:55 am:
This should be straightforward. Make up standards in ethanol or methanol, same for sample, use nonpolar capillary column.
![]()
![]()
![]()
![]()
By Ron on Wednesday, May 1, 2002 - 06:39 am:
There probably is an AOAC method for this analysis. I would use that as a starting point.
![]()
![]()
![]()
![]()
By Rodney George on Wednesday, May 1, 2002 - 11:35 am:
Supelco has an application note 55 for extraction of phthalates from water using SPE disks for quick concentrating and extraction of the esters.
Most methods indicate that a bonded MS capillary (Equity 1, for example) works very well for phthalates.
Rodney George
Supelco