Top of pagePrevious messageNext messageBottom of pageLink to this message  By berkerf on Wednesday, October 30, 2002 - 10:13 pm:

Hello,

I have a problem with column temperature selection. I want to analyze carbon dioxide and i dont know how to decide column oven temperature and detector temperature. I use a very old but working Varian 1400 aerograph with TCD detector. Column i use is Porapak Q.
If anybody can inform on settings of column oven temperature, detector temperature and carrier gas flowrate, i would be very pleased.

Thanks


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, October 31, 2002 - 05:37 am:

60°C oven

30cc/min carrier

detector temperature 120°C or higher