I want to build dual detector system : ECD and FID detectors.
Is there anything like automatic rotary valve - end column splitter- (similar to LC) which would enable me to close down automaticaly under program control either FID or ECD when that particular detector is not needed? I want to keep ECD as littile loaded as possible, when not in use. I would like to avoid manual column change.
I am looking forward to getting some general advice on dual detector systems, links and refcs. Thank you.
Tardus
![]()
![]()
![]()
![]()
By Bernd Mischke on Thursday, March 27, 2003 - 11:55 pm:
Plaese send me the type and configuration of Your GC. Never switch off the ECD gases, Oxygen destroy it few minutes later (the Ni63 foil) if the ECD is hot. A FID can switched off, air and H2 can be saved. Make up gas (recommended N2) should flow to clean the detector. What You can do automaticaly depends on the GC-system You have.