Kindly suggest purity analysis method of 1,2,4-Traizole on GC.
![]()
![]()
![]()
![]()
By maris on Saturday, April 19, 2003 - 09:09 pm:
Hi, anonymous, I have a hard copy. How do you want to recieve it? ( fax, mail ?)In case you want an e-attachement, it will takes me more time to prepare.
![]()
![]()
![]()
![]()
By akc on Sunday, April 20, 2003 - 11:23 pm:
Dear Maris
Thankyou very much for responding. You can fax the method on following no.
0129-5066911
Attn. AKC
Thanks once again
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, April 22, 2003 - 11:30 pm:
Dear Maris
Do not get any reply, pl. help
thanks
![]()
![]()
![]()
![]()
By maris on Thursday, April 24, 2003 - 12:22 am:
Dear Anonymous, unfortunately the fax doesn't works. I'll try to describe brifly:
Column: CAM from J&W (cat#115-2132), 30 m, 0,53mm,
1 micron
Pressure: 10 PSI
Injection volume 5 mcL
Injection temp - 250C
Detector FID - 275C
Temperature gradient: 140C for 2 min,
20C/min for 3 min, 2 min at 200C
Retention time of the main peak - ~ 5 min