Setup for C1 - C32 analysys

Chromatography Forum: GC Archives: Setup for C1 - C32 analysys
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, May 21, 2003 - 01:58 am:

Hi,

Can someone advise me on the best setup for
C1-C32 analysis. What's the minimum number of
columns and switches I need to get the best
results.

Thanks


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Rodney on Wednesday, May 21, 2003 - 07:30 am:

A single packed column will separate the C1-C5 effectively but it is suggested that you use a backflush to vent, single 10 port valve setup.

A capillary column will do the remainder C6-C32 through single dimension chromatography.

Email me for details.

Rodney George
Senior Research and Development Scientist
Gas Separations Research
Supelco
595 North Harrison Road
Bellefonte, PA 16823

814-359-5737 voice
814-359-5702 fax
rgeorge@sial.com


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, May 22, 2003 - 11:37 pm:

Thanks for your help Rodney