I use Chromosorb 102 in a packed glass column to determine low MW alcohols in beverages.It is very selective in that I only see aldehydes,ketones and alcohols eluting into the FID.Other compounds seem to be retained in the column/liner.
Is there a suitable capillary replacement I could try?
My samples contain up to 50% water!!!
Thanks in advance
![]()
![]()
![]()
![]()
By Anonymous on Wednesday, January 21, 2004 - 10:37 am:
Are you wanting to measure water? You are correct about other volatile components of beverages being retained on-column.
After heptane, other larger components get very broad on porous polymer packed columns.
What other components do you wish to measure?
Supelco's Equity 1701 capillary columns do a very nice job separating acetone, methanol, ethanol, isopropanol, n-propanol, and ethyl ether
and are durable to water injections.
Yes, I work for Supelco. Use our Tech service for more answers.