Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, February 16, 2004 - 01:54 am:

Hello!
Can anyone me suggest what would be still acceptable value for the tailing factor?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, February 16, 2004 - 05:39 am:

1.2


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Bernd Mischke on Monday, February 16, 2004 - 06:35 am:

Hello Anon,

don't be afraid but 1.5 is also acceptable in labs...


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, February 18, 2004 - 08:48 am:

Where does one obtain the tailing factor for a peak? Is it the same thing as the peak symmetry?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, February 18, 2004 - 09:53 am:

The values are reciprocal. TF of 1.5 = Sym of .67

They may be calculated from 10% of peak height or 5% of peak height, depending upon the source of the definition.