Chromatography Forum: GC Archives: Lactic acid 2
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, May 10, 2004 - 01:57 am:

Hi
Last week i received two answer about the disparition of lactic acid.I thank you those people but i have forgotten to tell you that when i inject lactic acid on its own , it is eluted .So why when it is mixed with other fatty acid it does no appear.We cleaned the liner and cut few mm of column.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, May 10, 2004 - 10:18 am:

Hi

Cut 30cm,and replace the liner;
or inject on column to see any differences...


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, May 11, 2004 - 01:57 am:

Try E-mailing Becky Wittrig at Restek. She has a lot of good answers on food analyses.

www.restekcorp.com

GC Dave


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, May 12, 2004 - 02:09 am:

thank you