Repeller 5973ms cleaning or changing?????

Chromatography Forum: GC Archives: Repeller 5973ms cleaning or changing?????
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, May 12, 2004 - 11:00 am:

Hi

i' ve cleaned my ms source today, but after cleaning,new filament installed,my repeller voltage is still at maximum.(35 V in Autotune)
It seems that cleaning this repeller bring nothing.
Does it make sense to buy a new one??????????,,,


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, May 13, 2004 - 08:45 am:

It is unlikely a new repeller will help. If the voltage is really getting to the repeller there is no way it should be maxed out. I would double check all connections to make sure everything is correct, then reclean, reassemble, then put in a service call if things still look bad.