Acrylic acid with GC?

Chromatography Forum: GC Archives: Acrylic acid with GC?
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 10, 2004 - 02:25 am:

Hi
I used to quantify acrylic acid with an FFAP column,but after 2/3 injections,the performance of the column is bad,tailing etc..
My parameters are on column inj. ffap 30*0.53*1.0
solvent Etoh oven programm70C 1min-10C/min-230C
Does anybody know?
Bye


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 10, 2004 - 06:05 am:

Polymerization of acrylic acid in the column at high temperature?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 10, 2004 - 12:37 pm:

What temperature his high enough for polymerisation?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 10, 2004 - 03:22 pm:

Acrylic acid, unless stabilized, will polymerize even at room temperature. At elevated temperatures the rate is much higher. You might want to try HPLC