DMF analysis by GC, in which medium can be analysed

Chromatography Forum: GC Archives: DMF analysis by GC, in which medium can be analysed
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, June 23, 2004 - 03:28 am:

Can any body suggest me about in which medium I can quantitate the Dimethyl formamide as a residual solvent by GC.I tried with DMSO only.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, June 23, 2004 - 03:48 am:

Dimethyl acetamide would be one solvent you could use. Be sure to run blanks so you know the amount of DMF in the DMAc.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By ingochrist on Tuesday, June 29, 2004 - 08:33 am:

I have done DMSO or DMF by headspace or SPDE (Solid Phase Dynamic Extraction). My detection limits were at the low ppm range with SPDE and high ppm with headspace.