Hello Friends'
Can any one explain to me what precuations to be taken in regenerating packed columns, espically molecular sieves column.
With Thanks
jeevan
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, June 29, 2004 - 08:27 am:
heat with carrier gas flowing
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, June 29, 2004 - 06:26 pm:
You need to install the column in a GC oven, detector connection is not recommended, buy may be used to monitor baseline. Set the column carrier flow to at least 30ml/min with helium. Raise the oven temperature to 300°C. Condition the colum for 2-4 hours. Do not over condition the column (too hot or too long), carbon monoxide peak may begin to tail slightly if you over condition the column.
Good Luck,
Omar