Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, June 29, 2004 - 05:55 am:

Hello Friends'

Can any one explain to me what precuations to be taken in regenerating packed columns, espically molecular sieves column.

With Thanks
jeevan


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, June 29, 2004 - 08:27 am:

heat with carrier gas flowing


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, June 29, 2004 - 06:26 pm:

You need to install the column in a GC oven, detector connection is not recommended, buy may be used to monitor baseline. Set the column carrier flow to at least 30ml/min with helium. Raise the oven temperature to 300°C. Condition the colum for 2-4 hours. Do not over condition the column (too hot or too long), carbon monoxide peak may begin to tail slightly if you over condition the column.

Good Luck,

Omar