Top of pagePrevious messageNext messageBottom of pageLink to this message  By linea on Friday, July 16, 2004 - 12:30 am:

Hi
I am new in Gc and try to undersatnd some points.What is baking out the column?Is it a kind of cleaning???
thank you


Top of pagePrevious messageNext messageBottom of pageLink to this message  By shabhari on Friday, July 16, 2004 - 01:38 am:

Linea

It's nothing but regenerating or more precisicely to tell,conditioning your column, inorder to improve it's performance.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By GC Dave on Friday, July 16, 2004 - 04:06 am:

Linea,
Its done by leaving the column at its isothermal temperature limit. This should be done for 30mins to 1hour initially and obviously with carrier gas flow.
Depending on your samples you may need more time.

Some non-volatiles should leave the column. Others will eventually build up requiring you to replace the column.

GC Dave


Top of pagePrevious messageNext messageBottom of pageLink to this message  By anonymous on Friday, July 16, 2004 - 11:18 am:

Linea,

Instructions for conditioning a column are always supplied with the column.