Detection of residual solvents in edible oil samples

Chromatography Forum: GC Archives: Detection of residual solvents in edible oil samples
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Ganesh ramachandran on Tuesday, August 10, 2004 - 12:07 am:

Can Anybody tell me how can we assay residual solvent in vegetable oils? Is head space analysis on GC reliable? Please help. Thanks.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By chromatographer1 on Tuesday, August 10, 2004 - 05:55 am:

Static headspace will be very accurate and reproducible. Add a portion of the oil to Dimethylacetamide (10µL to 50µL.... keep the volume consistent) and use 80°C heating with stirring for 10min. Inject amount needed to give adequate peak size..... 100 to 500µL .