My hydrogen peaks are small using helium as carrier. If I change to nitrogen as carrier I will probably loose my CO, CO2 and O2 peaks. Which carrier gas should I use?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Consumer Products Guy on Wednesday, August 25, 2004 - 08:40 am:

Try a custom mix of 8.5% hydrogen in helium as carrier; that's what we used for this. Hydrogen can be a negative peak, a positive peak, or small, depending on its concentration, that's why this custom blend is used. Look in literature, we didn't develop this ourselves.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By chromatographer1 on Monday, August 30, 2004 - 05:14 am:

Argon is another option for a quick fix.