Top of pagePrevious messageNext messageBottom of pageLink to this message  By Teresa Chong on Tuesday, November 16, 1999 - 05:17 am:

I need to know how sodium acetate can be analysed by a GC column. What kind of mobile phase should I be using?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By John Hinshaw on Friday, November 19, 1999 - 02:40 pm:

Teresa,

Please post your message on the LC forum. GC is not normally useful to analyse sodium acetate.

John Hinshaw


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Ulf Houlberg on Monday, December 27, 1999 - 05:13 am:

Hi Teresa
We are not normal !
Sodium acetate can easily be converted quantitatively into Methylacetate using Methanol and Sodiumhydrogensulfate and then analysed by headspace GC. The sample preparation take less than a minute pr sample and sub-ppm levels can be measured. We use the method rutinely to monitor acetate and formate in low levels. Revert if you want details.