Top of pagePrevious messageNext messageBottom of pageLink to this message  By Uberto on Saturday, April 13, 2002 - 11:05 pm:

Dear members,

Does someone know how to process/print "custom report" using DOLIST command as found in the MS Chemstation software? I recently build a custom report using such software but cannot process the data files in series using DOLIST. However, DOLIST can process/print summary, detail reports etc. of many data files at once. Please advise.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Richard Turcotte on Friday, May 31, 2002 - 10:44 am:

Hi!
You cannot choose to print custom report with Dolist.
I use dolist for calculate my result and after,i open custom report for print multiple file.

1-Dolist
2-Configure dolist option.....choose QUANT NO REPORT
3-Dolist
4-Use configured dolist option..choose quant no report
5-Choose your file and process
6-quantitate and open custom report
7-Edit template
8-file
9-multiple file select
10-choose your file calculated with dolist and process

Good luck!