Compatibility of DMSO with MS

Chromatography Forum: LC-MS & GC-MS Archives: Compatibility of DMSO with MS
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Pim on Monday, March 1, 2004 - 12:24 pm:

Developping an LC method, I observed that only DMSO as organic modifier lead to a sufficiently good method. I wonder if this solvent can well be used if I furhter want to couple LC to MS?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By MG on Monday, March 1, 2004 - 03:28 pm:

I don't know how DMSO behaves as far as getting good ionization. I've never used it in my mobile phase. But there's another issue that could present a problem. Some ion source designs use PEEK tubing as the transfer line between the ESI needle (at ~ 5kV) and a grounded union, for safety. I think PEEK is incompatible with DMSO. Replacing that particular length with stainless steel would short the power supply.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, March 1, 2004 - 03:48 pm:

DMSO is NOT volatile.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, March 1, 2004 - 04:21 pm:

DMSO bp (760) 189 degrees...


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, March 1, 2004 - 05:01 pm:

DMSO bp : 210-235 °C/760 mm Hg (lit.) degrees...


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, March 30, 2004 - 05:12 am:

So what's the best way to remove DMSO once inside the MS?