Disodium adduct formation in LCMS

Chromatography Forum: LC-MS & GC-MS Archives: Disodium adduct formation in LCMS
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, April 16, 2004 - 11:57 pm:

Dear All,

We have observed [M+1], [M + 23] and [M+45]for one of our samples in LCMS. Is the formation of Disodium adduct is possible?

Advance thanks


Top of pagePrevious messageNext messageBottom of pageLink to this message  By MG on Sunday, April 18, 2004 - 01:37 pm:

Yes, but if you put only 2 Na+ on there, your ion would be doubly charged, so you'd get (M+46)/2.

If your compound has an acidic functionality as well, you could have (M+2Na-H)+ or (M+45). I'm not sure if I've ever seen this particular one, but have seen combinations of adducts and losses. You should consider it along with other possibilities.