Top of pagePrevious messageNext messageBottom of pageLink to this message  By Vladimir Besada on Friday, September 29, 2000 - 06:48 am:

Hi:
Do someone have experience with the Michrom rp C30 columns? Any application on small biomolecules?
Thanks
Vladimir


Top of pagePrevious messageNext messageBottom of pageLink to this message  By frederic bell on Wednesday, October 25, 2000 - 03:07 am:

I do not have any but I plan to get some experience onthis column. I have a separation usually performed on a C18 column, however even with a 98 % aqueous buffer ratio in my mobile phase, the retention of some compounds is not sufficient. So I suppose using a C 30 column will help me to increase the retention of these compounds. Do you have any advice to give me in that concern ?

frederic


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, October 25, 2000 - 06:35 pm:

If you do not have enough retention with 98% buffer, why don't you go to 100%? If you are afraid about the hydrophobic collapse thing, you could use either as column without endcapping or a column with an embedded polar group.