Hi:
Do someone have experience with the Michrom rp C30 columns? Any application on small biomolecules?
Thanks
Vladimir
![]()
![]()
![]()
![]()
By frederic bell on Wednesday, October 25, 2000 - 03:07 am:
I do not have any but I plan to get some experience onthis column. I have a separation usually performed on a C18 column, however even with a 98 % aqueous buffer ratio in my mobile phase, the retention of some compounds is not sufficient. So I suppose using a C 30 column will help me to increase the retention of these compounds. Do you have any advice to give me in that concern ?
frederic
![]()
![]()
![]()
![]()
By Anonymous on Wednesday, October 25, 2000 - 06:35 pm:
If you do not have enough retention with 98% buffer, why don't you go to 100%? If you are afraid about the hydrophobic collapse thing, you could use either as column without endcapping or a column with an embedded polar group.