Volatile buffer for acid antibiotic

Chromatography Forum: LC Archives: Volatile buffer for acid antibiotic
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Jian LI on Monday, November 13, 2000 - 03:53 pm:

HI, All,

Does anyone know the proper volatile buffer for the acid antibiotic for the ESI-MS assay?

Many thanks in advance.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, November 16, 2000 - 08:39 am:

Acetate is one of the most common volatile buffers for MS. The pH range is about 4-6. Although not a buffer, TFA is used to adjust pH to 2-3.