I'm looking for HPLC column / instrument parameters / mobile phase that would work for the analysis of ammelides in Cyanuric acid. Please help.
![]()
![]()
![]()
![]()
By Anonymous on Friday, December 1, 2000 - 08:09 am:
Try this:
Column: NH2 reverse phase (25 x 0.46 cm)
MP: Acetonitrile : 5 mM pH 6.7 sodium phosphate buffer (78:22)
Flow: 1.0 ml/min.
UV: 214 nm
I've used this method in the past for melamine and related amino- & hydroxy triazines.
![]()
![]()
![]()
![]()
By Anonymous on Friday, December 1, 2000 - 12:38 pm:
Dear "NH2" -- I appreciate your help. Thanks
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, September 9, 2003 - 07:23 am:
SOMOS UNA EMPRESA DISTRIBUIDORA DE MATERIAL DE LABORATORIO EN ARAGON (SPAIN). NECESITAMOS OFERTA PARA UN STANDAR DE:
AMMELIDE PARA HPLC.
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, September 9, 2003 - 07:31 am:
somos una empresa de Distribución Material Labor. Quimico, estamos buscando un Standar de Ammelide
para HPLC. Rogamos nos oferten o nos digan como conseguirlo.
Gracias. Salvador diaz
![]()
![]()
![]()
![]()
By Einar Pontén - SeQuant AB on Tuesday, September 9, 2003 - 02:46 pm:
ZIC™-HILIC is a suitable column for ammelide (2,4-Dihydroxy-6-amino-1,3,5-triazine). Mobile phase: approx. 70% acetonitrile/30% 5 mM NH4Ac