Top of pagePrevious messageNext messageBottom of pageLink to this message  By Andrea on Tuesday, February 13, 2001 - 07:28 am:

I am trying to analyse some water samples that may be contaminated with MTBE. I need some tips on how to go about proving the presence or absence of MTBE using LC. Any ideas would be appreciated.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, February 13, 2001 - 08:29 am:

Why LC, I would think that GC (GC/MS for positive compound ID) would be the way to go here. By the way, most methods can not determine the absolute absence of a given compound. They can determine that amounts found are more than the detection limit, but not less. This is a very important distinction to make when doing the kind of environmental contamination analysis that you are attempting.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, February 13, 2001 - 08:33 am:

For the detection of trace MTBE in gasoline contaminated soils or water, I recommend EPA Method 8260, which utilizes GCMS. GCMS gives unequivocal results for MTBE due to its unique mass spectrum, which is easily distinguishable from potentially interfering matrix components.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Tom M on Tuesday, February 13, 2001 - 08:36 am:

I too think this is a natural for cap. GC. There is an EPA method for volatile organics, I think it is 8021A that uses a PID. You could also use a CP-SIL5 CB column with a FID to detect volitle organics. If you want details on either of these methods you can email me. Goo Luck.