I found a literature reference to a separation that I would like to perform. The mobile phase used in the reference is a 25 mM phosphate buffer/acetonitrile gradient. Could anyone give me a procedure for preparing the 25 mM phosphate buffer? Also from the way it was stated in the reference, I assume the buffer is only added to the water and not to the acetonitrile. Is this correct? Thanks in advance for your assistance.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By I.D. on Wednesday, March 14, 2001 - 10:45 am:

Try this:

http://www.bi.umist.ac.uk/users/mjfrbn/buffers/makebuf.asp


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Merlin on Wednesday, March 14, 2001 - 04:33 pm:

Dissolve 25 mmole (FW x 0.025) phosphate in 1 L water. Adjust pH if necessary. If the method has pure MeCN as the B mobile phase, there definitely is no phosphate in it as it won't dissolve.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By H W Muller on Wednesday, March 14, 2001 - 08:59 pm:

Check"Alcohols and pH" February 6, 2001