Comparison of hydrophobicity of c8, c18 and PS-DVS

Chromatography Forum: LC Archives: Comparison of hydrophobicity of c8, c18 and PS-DVS
Top of pagePrevious messageNext messageBottom of pageLink to this message  By madhav on Thursday, May 3, 2001 - 09:58 pm:

Hi!

Just wondering if we can compare hydrophobicity of c8 and c18 Vs polystyrene-divinilbenzene.
Can we extrapolate RPC data from c8 or c18 to PS-DVB to isolate drug molecules from impurities?

Your inputs would be of great help.

thanks in advance.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Uwe Neue on Friday, May 4, 2001 - 04:07 pm:

Both work in a similar way, when used with reversed-phase mobile phases. However, for a prediction of retention for the preparative isolation I think you need a fairly good data set for calibration. In addition, the selectivity is guaranteed to be different. If C18 resilves the impurities, I would not bet that PS-DVB would do as well.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, May 10, 2001 - 01:10 am:

PS-DVB has significant differences in selectivity. The PS-DVB has a greater affinity for aromatics compared with standard Reversed Phase columns.