Top of pagePrevious messageNext messageBottom of pageLink to this message  By chris knutsen on Friday, June 1, 2001 - 04:55 am:

I am looking for a method to detect 10 ppb levels
of N-nitrosodiethanolamine in diethanolamine. I know that this is very low and will probably entail a concentration step, but if anyone has any ideas I would really, really, really appreciate it.

Thank you in advance and everyone have a great weekend

Chris


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, June 1, 2001 - 12:33 pm:

chemiluminescence detector? (does anyone still make one?)


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Joe Runkle on Friday, June 1, 2001 - 12:46 pm:

Chris,

I did some work on NDELA in cosmetics on LC/MS with APCI ionization. I was working at ~150 ppb and had decent S/N by monitoring the m+1 SIM at 135 m/z. 10 uL injection, 5 reps gave a mean response of ~80000 area counts (S/N ~150) RSD of ~10%. If you want method specifics, or if I can be of help, feel free to email.

Have a good weekend.

Joe