Top of pagePrevious messageNext messageBottom of pageLink to this message  By warren on Tuesday, July 6, 1999 - 03:27 pm:

Can anyone recommend a good type of solvent bottle, possibly air/gas tight, that can contain lines for sparging and flow for HPLCs. possible evaporation problem.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By hinsbarlab on Wednesday, July 7, 1999 - 09:11 am:

You shouldn't use an airtight bottle. Removal of the mobile phase will cause a partial vacuum as the bottle is emptied and probably cause flow problems. Alltech sells a pressurized system (pg. 611, Catalog #450) for use with helium sparging that works very well. A word of caution - when using pressurized systems be sure that an overpressure safety valve is incorporated into the apparatus.

Michael Hinsberg
Hinsbar Laboratories, Inc.
www.hinsbarlabs.com


Top of pagePrevious messageNext messageBottom of pageLink to this message  By warren on Wednesday, July 7, 1999 - 01:48 pm:

Thanks for the info