Can anyone recommend a good type of solvent bottle, possibly air/gas tight, that can contain lines for sparging and flow for HPLCs. possible evaporation problem.
![]()
![]()
![]()
![]()
By hinsbarlab on Wednesday, July 7, 1999 - 09:11 am:
You shouldn't use an airtight bottle. Removal of the mobile phase will cause a partial vacuum as the bottle is emptied and probably cause flow problems. Alltech sells a pressurized system (pg. 611, Catalog #450) for use with helium sparging that works very well. A word of caution - when using pressurized systems be sure that an overpressure safety valve is incorporated into the apparatus.
Michael Hinsberg
Hinsbar Laboratories, Inc.
www.hinsbarlabs.com
![]()
![]()
![]()
![]()
By warren on Wednesday, July 7, 1999 - 01:48 pm:
Thanks for the info
Posting is currently disabled in this topic. Contact your discussion moderator for more information.