Top of pagePrevious messageNext messageBottom of pageLink to this message  By Fabiano on Friday, April 19, 2002 - 07:23 am:

Hello,

Anyone already try to connect a integrator (like Shimadzu CR6-a)into a computer?
some models like from HP have a "computer port" behind,seens to be a 2 wire cable ,no com1 or com2 port but even this kind are too strange for me. the detector is a shimadzu UV-Vis .

Regards

Fabiano


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, April 19, 2002 - 10:24 am:

If you are trying to get the signal to a PC, the best thing to do is to run the signal from the detector through an A to D board and into a data collection package on the PC. Some integrators have a floppy drive, and software is available to transfer files from the integrator to a PC.