Hello,
I´d like to know where i could find information about kh2po4 solubility in organic phases; i have to study the influence of pH in my mobile phase and i suspect that 5mM kh2po4 is not totally soluble in 45%AcN;I´m not sure, because i can see like threads and is not the usual way of precipitation of phosfates..
I have to study from ph 3 to ph 10.It is possible that solubility decreases increasing ph?
Thanks
![]()
![]()
![]()
![]()
By Anonymous on Thursday, August 1, 2002 - 04:05 pm:
5 mM KH2PO4 is soluble in 45% MeCN. No problem!
![]()
![]()
![]()
![]()
By Celia on Wednesday, August 7, 2002 - 07:00 am: