HPLC method of DONEPEZIL (ARICEPT)

Chromatography Forum: LC Archives: HPLC method of DONEPEZIL (ARICEPT)
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, December 10, 2002 - 06:38 am:

Hi to all,

does someone know about an HPLC analytical method [using UV-VIS detector]of DONEPEZIL(ARICEPT)?

Thanks in advance


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Jan on Friday, January 3, 2003 - 03:24 am:

Hello Anonymous,
Pfizer uses Inertsil ODS-2 5 µm 150 x 4.6 at 35°C. The mobile phase is prepared as follows: 2.5 g sodium decanesulfonate in 650 ml water. Add 1 ml of 70% perchloric acid and 350 ml acetonitrile. Adjust apparent pH to 2.5. Flow 1.4 ml/min. 271 nm. Inject 20 µl of 0.2 mg/ml donezepil HCl in dilution solvent (0.1 N HCl - Methanol 1:3). Donezepil elutes after appr. 12 min.
Hope it works
Jan


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, January 3, 2003 - 05:39 am:

u search in chemweb.com ,merhods available.all the best.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, January 3, 2003 - 05:42 am:

Adjust pH after adding acetonitrile? I thought, that we should measure pH in aqueous phase before mixing with organic solvent.
Could you explain this, for I’m new in HPLC, and getting confused.

Have a great 2003!!!


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Jan on Friday, January 3, 2003 - 06:24 am:

I admit adjusting pH before adding organic sovent is more staightforward. Anyway, I believe this particular method was developed by adjusting pH AFTER adding acetonitrile (that's why I wrote apparent pH). John Dolan had a very interesting column on this in the November or December 2002 issue of LC GC (a monthly published FREE journal, and available online)!
http://www.chromatographyonline.com/lcgc/data/articlestandard/lcgc/462002/38298/article.pdf


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, January 3, 2003 - 06:42 am:

Jan:
I work in the industry. Method details are company property. Giving confidential info to third party could cause trouble. I appreciate your kindness but you might be more careful in the future.
Best regards.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, March 9, 2004 - 06:14 am:

hplc method for purity determination of moxifloxacin and gatifloxacin along with their acid and esters ( ie., with out their side chains re., nonane and 1 methyl piperizine)


Top of pagePrevious messageNext messageBottom of pageLink to this message  By aniket a naik on Monday, June 21, 2004 - 12:31 am:

does anyone know TLC method for analysis of donepezil.
Thank you
aniket