Guard cartridge/column for microbore columns?

Chromatography Forum: LC Archives: Guard cartridge/column for microbore columns?
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Ann on Wednesday, April 9, 2003 - 10:16 am:

Dear Chromatographers

I am currently developing a HPLC assay for the analysis of a drug in whole human blood (sample clean-up is performed using solvent extraction). I am using a 1mm microbore column (C1) in order to enhance the sensitivity of the assay. I usually use the Phenomenex "SecurityGuard" cartridge systems for column protection but the smallest cartridges are only suitable for columns with a 2mm diameter (peak shape is severely compromised if I attempt to use these). Hence, I am currently using a combination of a Phenomenex "Krudcatcher" (a disposable in-line filter, 0.5u) and a standard stainless steel in-line filter (also 0.5u).

Does anybody know of a manufacturer who produces guard cartridges or columns suitable for use with microbore (1mm) columns?

Many thanks for any suggestions :o)

Ann


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Einar Pontén - SeQuant AB on Wednesday, April 9, 2003 - 02:41 pm:

Ann, SeQuant has 1 x 10 mm stand-alone guard columns. Send us an e-mail and we will try to advice you.