Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, May 5, 2003 - 07:05 am:

Can anyone please supply me with a simple CE application. I am fairly new to this technique, and would like to try out a simple application that is sure to work.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By bill tindall on Monday, May 5, 2003 - 04:57 pm:

any aromatic acid, for example benzoic.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, May 6, 2003 - 12:44 am:

what buffer can I use?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Jan on Tuesday, May 6, 2003 - 02:45 am:

20 mM sodiumtetraborate (pH about 9) is a good way to start. will not produce to much current as opposed to phosphate buffers. You can go upto 20 or 30 kV if the capillary ID is equal to or less than 75 µm. If you are new to this technique and want to produce a nice electropherogram, degas a few ml of diet coke (no sugar !!!) and inject. You'll find caffeine, benzoic acid and aspartam (amongst other smaller peaks).
Good luck
Jan


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, May 7, 2003 - 12:53 am:

Thanks Jan