Looking for good HPLC procedure for Alendronate Sodium in Tablets

Chromatography Forum: LC Archives: Looking for good HPLC procedure for Alendronate Sodium in Tablets
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, June 3, 2003 - 02:24 am:

The information available for analysis of Alendronate sodium in tablet formulations is by using the Refractive index Detector.

I am looking for a method without using the above mentioned, instead useing a Uv/Vis detector.

Please help me in this regards

Dr.R.M.RAO
My e-mail as follows:

drrayapudi@yahoo.com


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Jan on Tuesday, June 3, 2003 - 02:35 am:

Sodium Alendronate has no real chromophores. If you still want to use UV/VIS detection, you can
a) try low wavelength
b) derivatise (pre or post column)
c) use indirect detection

Another way to go is IC (have no experience myself, but believe it should work)