I haven't got any pressure at all in my system, though i have checked the usual causes (leaks, pump, etc).
wat do you think may be causing it?
thanks in advance
![]()
![]()
![]()
![]()
By read on Monday, June 23, 2003 - 06:02 pm:
Do you have flow from the pump? Bleed out all air
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, June 24, 2003 - 04:22 am:
Make sure there's a good flow up to the inlet check valve. Disconnect the tubing at that juncture. I bet the check valve is "frozen" open.