Help! Waters 2487 Display "Fatal error"

Chromatography Forum: LC Archives: Help! Waters 2487 Display "Fatal error"
Top of pagePrevious messageNext messageBottom of pageLink to this message  By wanghuai on Thursday, June 26, 2003 - 08:00 am:

Help ! The 2487 Detector display the word , What's wrong?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, June 26, 2003 - 08:30 am:

Did you look in the manual on page 175?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Friday, June 27, 2003 - 06:11 am:

yes, but the fatal error persists.

If not contact Waters technical servce, can solve the problem?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Martin on Monday, June 30, 2003 - 12:56 am:

Please give a description of situation in detail und perhaps the full error code or message.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By wanghuai on Tuesday, July 1, 2003 - 06:56 am:

startup
first: Lamp data not found
two: System not calibrated
(lamp warmming 5 min)
three:Method not found

than display "error".


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, July 1, 2003 - 12:19 pm:

Your battery is shot. You need to order a new one. After you install the new battery, you will need to re-calibrate which is not hard to do.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, July 2, 2003 - 07:19 am:

Battery may be Panasonic CR2032, 3V. There is no need to buy it from Waters.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By wanghuai on Friday, July 4, 2003 - 10:29 pm:

thanks!


Top of pagePrevious messageNext messageBottom of pageLink to this message  By wanghuai on Sunday, July 6, 2003 - 11:50 pm:

i have replaced the battery, but (two: System not calibrated )

How to re-calibrate the system?

tell me the detals step by step,please.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Monday, July 7, 2003 - 02:16 am:

1. flush the cell with AcN,MeOH or water (UV transparent solvent). Remove cuvette (if inserted).
2. press shift and 3 (calibrate)
3. wait for a while, the system show you results.

Sometimes you will have to press "ENTER"...

Anyway, you should have operator's guide. The calibration is described on page 3-28.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By wanghuai on Monday, July 7, 2003 - 06:55 am:

thanks !!!

i will try it