Chromatography Forum: LC Archives: Mobile
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, August 6, 2003 - 10:48 pm:

what is the necessity to dissolve the sample in Mobile phase (as in different pharmacopie)?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, August 6, 2003 - 11:45 pm:

to prevents mixing effects between mobile and injection solvent (peak broadening, solvent peaks..)


Top of pagePrevious messageNext messageBottom of pageLink to this message  By M_Gardner on Thursday, August 7, 2003 - 06:32 am:

A solvent with stronger eluting strength than your mobile phase will act like a strong mobile phase briefly when it hits the head of your column, causing the problems mentioned by anon # 2. Your sample doesn't have to be dissolved in your starting mobile phase, but if not, it should be dissolved in a *weaker* solvent than your starting mobile phase.