Signal to Noise Ratio

Chromatography Forum: LC Archives: Signal to Noise Ratio
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Aryo on Monday, September 27, 1999 - 11:43 am:

How do you determine the S/N? Do you measure the Noise in presence of analyte of interest or independently (by injecting blank)?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By B.Buglio on Monday, September 27, 1999 - 08:59 pm:

Run the analyte at a low concentration and high
attenuation. Measure the peak height . Select a
stable segment of baseline. Draw parallel lines to
encompass the + and - deflections. Measure the
distance betwwen the parallel lines. The S/N ratio
at the given concentration is the analyte peak
height /noise peak h