Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, December 2, 2003 - 06:35 am:

I analyse samples with HP1100 HPLC,the software is HP1100 chemstation,I want to report tailing factor in my result report,how can i do?.
i use the report style is performance.
resolution ,plate counts are all in report except tailing factor.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By mahmoud on Tuesday, December 2, 2003 - 10:26 am:

Just go the report and use extended performance and make sure to check taling factor option. In the menu "Method and run" go to sequence and click on extended statistics to have tailing factor option. Hope that works,
Mahmoud


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, December 3, 2003 - 06:25 am:

check the calculations manually cause there is a bug in the chemstation calculations of the usp tailing factor in case of a sloping baseline(atleast in A.08, maybe in 09 as well).


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, December 17, 2003 - 04:24 am:

Thanks of you,I can take it.