Detection of sucralose - comparison of columns

Chromatography Forum: LC Archives: Detection of sucralose - comparison of columns
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, December 18, 2003 - 02:12 am:

I am developing a method for sucralose using RP-HPLC. Literature tells me to use C18 with Water/Organic. I have both BDS-C18 and Discovery RP-Amide C16.

When I test both columns under 70/30 Water/MeCN the Discovery C16 amide shows a lot more peak area for sucralose (factor of 50-100) at 190nm.

I have also tried with a Bonus-RP Amide and I get low area too.

Has anyone an explanation?

Thanks in advance for your thoughts


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Thursday, December 18, 2003 - 02:56 am:

Sorry Chromatographers!!

I have solved my problem - I had used TFA instead of hplc water in some of my work.

Poor troubleshooting on my behalf