Bed volume vs. injection volume?

Chromatography Forum: LC Archives: Bed volume vs. injection volume?
Top of pagePrevious messageNext messageBottom of pageLink to this message  By steve on Thursday, February 19, 2004 - 07:38 pm:

where can i find some guidelines on the proper injection volumes for columns of size X? i saw something recently that said the recommended injection volume for a 75um ID capillary column was 0.05 uL. how can one load a sample with a volume that small?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Uwe Neue on Friday, February 20, 2004 - 03:27 pm:

It depends on how you make up your sample. If your sample is dissolved in a compatible solvent, you may be able to inject a sample volume of 1000 column volumes.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, February 24, 2004 - 04:53 am:

very small volumes can be loaded by doing a split of the injection-flow.