Aromatic amines in methanol

Chromatography Forum: LC Archives: Aromatic amines in methanol
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Tuesday, February 24, 2004 - 06:21 am:

I've got 2 questions. Firstly I've been analysing aromatic amines using a strong cation exchange on silica (Luna SCX) and get good retention (mp 20mM KH2PO4 85% ACN 15%). I get a nice separation but I have no feel for how robust these SCX methods are. Has anyone stressed one of these columns with a thousand injections? and has it been robust?
The other thing I could try is derivatisation. A lot of procedures use ortho-phthalaldehyde (try saying that after a few beers!). Given that the amines are in Methanol is that the easiest way to go?
Regards
John


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Consumer Products Guy on Tuesday, February 24, 2004 - 06:39 am:

I wouldn't bother with derivatization as your compounds are already aromatics (UV detection) and you state that you're getting good separation; why add extra step(s)? I'd derivatize for non-aromatic amines.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, February 25, 2004 - 12:26 am:

One of the aromatic amines has an acid (SO3H)group as well and appears to be too polar to be retained in any system I've tried. I may have to derivatize because of that one.
regards
John


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Uwe Neue on Wednesday, February 25, 2004 - 03:27 pm:

You can potentially use silica itself as the ion-exchanger. It is a weak ion-exchanger, of course, but since there is nothing on the surface, nothing can wash off. It is commonly used for HILIC applications with less than 50% water.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Andreas Neumaier on Wednesday, March 3, 2004 - 01:38 am:

Another way to seperate aromate amines is to use a c8/18 with high conc of water (85 to 95%) at low pH (0,5 to 1% of 0,5M sulfuric acid).
When seperation of the acidic one from the others fails with this conditions use a YMC Pro c4/8/18 because they show some very interesting increase of retention for aromates with -SO3H groups under such conditions.
Maybe you have to run a (fast) gradient to seperate all the compounds in pleasent time.