Top of pagePrevious messageNext messageBottom of pageLink to this message  By Brent on Friday, March 12, 2004 - 01:45 pm:

Has anyone been able to separate these under conditions suitable for LC-MS?


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Chris Pohl on Friday, March 12, 2004 - 04:07 pm:

I haven't tried separating all three of these acids but I'm fairly certain that they can be separated at high pH on an IonPac AS11 column from Dionex. When used in conjunction with a suppressor (such as the ASRS Ultra II), this separation can be used in conjunction with LC-MS. If you're interested, I can obtain a sample of mannuronic acid to verify this separation capability.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By AllsepTech on Friday, March 12, 2004 - 08:25 pm:

Brent,
We are developing a method for sugars and their respective acids and will update you shortly. Method will envolve Primesep B or Primesep B2 column and LC/MS compatible mobile phase (ACN, water and volatile acid or formate) with our Primesep B or Primesep B2 column). Check our website for updates www.primesep.com


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Brent on Thursday, March 18, 2004 - 08:02 am:

Thanks for the replies, my email is bboudreaux@delsite.com