LOD and LOQ according to Ph. Eur.

Chromatography Forum: LC Archives: LOD and LOQ according to Ph. Eur.
Top of pagePrevious messageNext messageBottom of pageLink to this message  By rr on Tuesday, July 13, 2004 - 11:58 pm:

According to European Pharmacopoeia signal to noise ratio is calculated as 2H/h, where H=height of the signal and h=noise level.
I do not understand why a multiplication factor of 2 is used.
According to the mentioned calculation in EP, the ratio of height of peak to noise has to be 1.5, to meet the Limit of Detection criteria of S/N=3


Top of pagePrevious messageNext messageBottom of pageLink to this message  By ps on Friday, July 16, 2004 - 10:30 pm:

I too have the same question and this probably need to ask Ph.eur authorities itself.

But, What I presume is that the noise considered is the half of the noise of 'p to p'