Top of pagePrevious messageNext messageBottom of pageLink to this message  By Olof J on Thursday, November 4, 1999 - 11:46 pm:

Hello all,
I will change from computor to manually handling of my SP-HPLC. I have no idea how I will conect the instrument so I get the Pump to stop when the last injection is done.
To day the labnet cable is still there and I have tryed to conect the "pump stop" out put on the autosampler to the "pumpstop" input on the pump. This works but then the pump is locked and I have to turn it of to overcome this.O


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Eva S on Friday, January 21, 2000 - 12:49 am:

You can use the queue on the pump to adress how many starts it will do on each method, there you can create a shutdown method with flow rate zero.This works for me, but I usually leave a low flow on.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By John Butler on Friday, February 4, 2000 - 06:40 am:

There is also a key stoke on the autosampler that turns off the pump stop manually. What version autosampler do you have? its different between the 8880 and 8780 series.

John