I see a message that there is a simple chromatographic method with a solution hexane sulfonic acid for analysing quaternary ammonium chlorides!
![]()
![]()
![]()
![]()
By SIELC_Tech on Tuesday, August 24, 2004 - 06:36 am:
Dear Isabel,
Please check the following links for the analysis of quaternary amines. You do not have to use ion-pairing reagent (hexane sulfonic acid)with our Primesep columns. Two main problems with quaternary amines is lack of retention on RP column and poor peak shape.
http://allsep.com/makeChr.php?chr=Chr_047
http://allsep.com/makeChr.php?chr=Chr_043
http://allsep.com/makeChr.php?chr=Chr_007
Here is the comparisson for regular RP and Primesep column:
http://allsep.com/makeCmp.php?cmp=Cmp_080
Please contact us if you need help in developing a new method.
![]()
![]()
![]()
![]()
By isabel on Tuesday, August 24, 2004 - 06:43 am:
Thank's a lot but i want a specify uv method for chloromequat choline chloride and choline chloride
![]()
![]()
![]()
![]()
By SIELC_Tech on Tuesday, August 24, 2004 - 07:50 am:
Cholines in general are not UV acive (accept derivatives like acetyl choline, carbachol, etc.). You either have to do derivatization or use inderect detection. Other approaches include LC/MS and ELSD detection techniques.
![]()
![]()
![]()
![]()
By A.N. Onimis on Wednesday, August 25, 2004 - 07:35 pm:
Go here: http://www2.waters.com/watprod.nsf/Newdocs/WA40482
You may need to register.
![]()
![]()
![]()
![]()
By Chris Pohl on Thursday, August 26, 2004 - 06:27 pm:
Compounds such as choline and acetyl choline are easily detected without derivatization by making use of conductivity detection http://www1.dionex.com/en-us/webdocs/application/industry/lifesci/ic/AN70_V15.pdf