Export data from PC1000.

Chromatography Forum: LC Archives: Export data from PC1000.
Top of pagePrevious messageNext messageBottom of pageLink to this message  By Anonymous on Wednesday, January 19, 2000 - 01:59 am:

Is it possible to export data from PC1000 to AIA format? Thank you.


Top of pagePrevious messageNext messageBottom of pageLink to this message  By Tom Jupille on Friday, January 21, 2000 - 12:00 am:

I haven't checked recently, but my notes from a few years back give the following:

X. Thermo Separations

Procedure for converting a chromatogram in the ThermSep. PC1000
software to a CDF file which can be read by (e.g.) Chrom Merge.

A. Open the Data Maintenance module
B. From the menu bar, select Conversion / Export to CDF format...
1. Select the file (*.RES) to export from the scroll box
a. If appropriate, select the injection # and wavelength
2. Input the name and path for the *.CDF file (keep in mind the
8+3 DOS/Windows 3.x limitation (Windows 95 will "fix" this)

Hope this helps!

-- Tom Jupille / LC Resources Inc.