In case of related substances detemination, is it necessary to take the moisture content of working standard before use for hygroscopic material and then use the corrected potency for calculations?
![]()
![]()
![]()
![]()
By Anonymous on Wednesday, February 16, 2000 - 10:37 am:
When using a reference standard, the standard should be well characterized and the purity known. For example, hypothetically, if your standard is "100% pure" in the sense that there are no other compounds except it contains 1.5% water, the purity of your standard is 100%-1.5% = 98.5%. Your calculation must take this into account.
![]()
![]()
![]()
![]()
By Anonymous on Wednesday, May 26, 2004 - 04:38 am:
do we need to consider any salt conversion factor in calculating the related substances