I'm using a Zorbax C8 cartridge column with a guard column attached. Mobile phase is 100% ACN.
What is the best sample to evaluate t0, the column dead time?
I was using 1 mM uracil, but the "unretained" peak seems to tail, and there is also a concentration dependence on the Rt, implying nonlinear chromatography. Would NaNO3 be a better sample to measure t0?
![]()
![]()
![]()
![]()
By Anonymous on Tuesday, February 29, 2000 - 07:19 pm:
Spike your sample with a touch af acetone. It is unretained and is easily detected by UV (unlike NaNO3). Based on your column and mobil phase, you could pick from a large variety of compounds because little is going to be retained.
Posting is currently disabled in this topic. Contact your discussion moderator for more information.